Zeiss IOL Master V5 (Pre-Owned)
The Zeiss IOL Master V5 optical biometer offers fast and precise measurements of ocular characteristics necessary for intraocular lens (IOL) power calculations. This non-contact instrument measures of axial length, corneal curvature, anterior chamber depth, and “white-to-white”. The easy-to-use interface clearly guides the operator through the measuring modes and can be performed automatically with the press of a button.
With the IOL Master, the axial lengths and radii of the two eyes are compared and their plausibility is checked offering more reliability, especially with problematic eyes. The system automatically calculates lens power to be implanted using scientifically recognized biometry formulas and then a lens model selected from a clear, integrated database.
The pre-owned IOL Master V5 from EMS includes a warranty and lifetime phone support. Used Zeiss diagnostic equipment is refurbished and recalibrated by certified, manufacturer-trained technicians.
Available:
- Pre-Owned
- Measuring ranges
- Axial length 14 – 40 mm
- Corneal radii 5 – 10 mm
- Depth of anterior chamber 1.5 – 6.5 mm
- White-to-white 8 – 16 mm
- Scaling of display
- Axial length 0.01 mm
- Corneal radii 0.01 mm
- Depth of anterior chamber 0.01 mm
- White-to-white 0.1 mm
- Formulas for IOL calculation
- SRK® II, SRK®/T, Holladay, Hoffer Q, Haigis
- Clinical history and contact lens fitting method for calculation of the corneal refractive power subsequent to refractive corneal surgery
- Haigis-L for IOL calculation for eyes after myopic/hyperopic LASIK/PRK/LASEK
- Calculation of phakic anterior and posterior chamber implants
- Optimization of IOL constants Interfaces
- Data interface for office management systems
- Data export to CD-RW or USB storage media
- Provision of data for Holladay IOL Consultant and HIC.SOAP Pro
- Ethernet port for network connection and network printer
- Compliance with HIPAA directive (USA only)in regard to user authentification and identification
Dimensions (footprint): 390 mm x 300 mm
Height: max. 610 mm (headrest)
Weight: approx. 18 kg
Rated voltage; frequency basic device: 100 to 240 V AC (10 %); 50/60 Hz
Power consumption basic unit: 90VA
Earth conductor: Device should be connected only to sockets with an intact earth conductor
Protection class: I
Protection type: IP 20
Device type: B (DIN EN 60601-1)
Power isolation transformer
Rated voltage; frequency power isolation transformer: 100 to 127 V AC (10 %); 60 Hz or 220 to 240 V AC (10 %); 50 Hz
Power consumption power isolation transformer: max. 115 VA (total power consumption of connected external devices)
Power isolation transformer fuses: 2 x T3.15 A H 250 V 5x20 IEC 60127 for 100 to 127 V AC
2 x T1.6 A H 250 V 5x20 IEC 60127 for 220 to 240 V AC
Environmental conditions for intended use
Temperature: 10 to 35 C
Relative humidity: 30 to 75 %, no condensation
Air pressure: 800 to 1060 hPa
Storage environment
Temperature: -10 to +55 °C
Relative humidity: 10 to 95 %, no condensation
Air pressure: 700 to 1060 hPa
Ambient conditions for storage and transport in original packaging
Temperature: -40 to +70 °C
Relative humidity: 10 to 95 %, no condensation
Air pressure: 500 to 1060 hPa
Measuring range
Axial length
Area: 14 to 40mm
Resolution of display: 0.01mm
Keratometer
Area: 5 to 10mm
Resolution of display: 0.01mm
Anterior chamber depth
Area: 1.5 to 6.5 mm
Resolution of display: 0.01mm
White-to-white (optional)
Area: 8 to 16mm
Resolution of display: 0.1mm
As a division of the Advancing Eyecare Alliance, Enhanced Medical Services is proud to provide financing options for all of your equipment needs.
Through Advancing Eyecare Finance, take advantage of multiple finance structures designed to meet your budget and ownership goals.
Request A Quote
Our team is dedicated to helping you find the right equipment to fit your needs and budget. Simply fill out the form below and one of our experienced staff will get back with you promptly with a quote. If you don't see what you're looking for, be sure to ask as new inventory is arriving daily.